   SERIES EL CONTROL CHARACTERISTICS
SPECIFICATION RATING/DESCRIPTION CONDITION
USB Interface Uses IEEE 488.2 and SCPI
commands and queries
Requires no cost downloadable driver
RS 232 Interface Uses IEEE 488.2 and SCPI
commands and queries
Baud Rate: 38400; Parity: None
Data Bits: 8; Stop Bits: 1; Flow control: None
GPIB Interface Uses IEEE 488.2 and SCPI
commands and queries
Optional card provides access to GPIB interface.
Add suffix "G" to model number when ordering.
Ethernet Interface Uses IEEE 488.2 and SCPI
commands and queries
Optional card provides access to Ethernet interface.
Add suffix "E" to model number when ordering.

   TABLE OF SAMPLE SOFTWARE COMMANDS
RUNCTION PURPOSE COMMAND EXAMPLE
OPERATION MANAGEMENT
Mode Selection of operating mode MODE CURRent, MODE VOLTage,
MODE POwer, MODE RESistance,
MODE CONDuctance, MODE SHOrt, MODE OFF
Load ON/Load OFF Engage/Disengage load with device under test INPut 1=ON, INPut 0=OFF
Set Parameter Value Set current, voltage, power,
resistance, conductance values
CURRent xxxx, VOLTage xxxx, POWer xxxx, RESistance xxxx, CONDuctance xxxx
Measure Measure Voltage, Current, Power MEASure:Current?, MEASure:VOLTage?
Protection Allows setting maximum parameter values
to protect the device under test as needed
as well as minimum voltage
VOLTage:PROTection:UNDer xxxx
VOLTage:PROTection:OVEr xxxx
Damping Selection Different damping settings to
allow optimizing the user's test
SYSTem:PFModule:DAMPing x,
SYSTem:DAMPing y
where x is a number from 0 to 5 (min - max)
where y is a number from 0 to 7 (min - max)
Transient Operation To create a transient input change to a new level for a
period of time and then return to the original setting.
Can also be used to create multiple transients.
Curr:TRAN 10,.1,5,8 create 8-5 second transients.
Each transient is .1 seconds at 10 Amperes
then return to original set point for 4.9 seconds.
Ramps To change the fast response time of the
EL (under 50 microseconds) to a period of
100 microseconds to 10 seconds.
SYST:RAMP.1
This is an absolute time of change and
not in amperes per second to reduce calculation
load on the test software.
SYSTEM MANAGEMENT
Power FET Modules Installed List power FET modules installed in EL SYSTem:PFModule:LIST?
Display Power FET Status Display FET temperatures, currents
and fuse status
SYSTem:PFModule:STATus?
External Analog Control On/Off Turn external analog input OFF or enable one of two inputs:
a high-speed hazardous input or isolated SELV
SYSTem:EXTernal 0 (disabled)
SYSTem:EXTernal 1 (hazardous enabled)
SYSTem:EXTernal 2 (isolated SELV enabled)

NOTE: System responds to short forms of all commands and queries. See instruction manual for details.